Respironics V60 Non-invasive ventilator
Description
Respironics V60
Non-invasive ventilator
Specifications Show less
- General
-
Oxygen inlet pressure range 276 to 600 kPa (40 to 87 psig)Weight with optional battery 10.9 kg (24 lb)Weight without optional battery 10.0 kg (22 lb)
Dimensions 33.7 cm H x 39.4 cm W x 42.9 cm D (13.3"H x 15.5" W x 16.5" D)
- Patient types
-
Suitability Pediatric (≥ 20 kg) and Adult
- Electrical
-
AC voltage 100 to 240 VACAC frequency 50 to 60 HzAC power 300 V·A
Battery operating time 6 hours in normal conditions
- Settings
-
C-Flex Off, 1 to 3IPAP 4 to 40 cm H₂OCPAP 4 to 25 cm H₂OEPAP 4 to 25 cm H₂OMax P (PPV maximum pressure limit) 5 to 40 cm H₂OIPAP 4 to 40 cm H₂OMax P (AVAPS maximum IPAP) 6 to 40 cm H₂OInspiratory time 0.30 to 3.00 secMax P (AVAPS maximum IPAP) 6 to 40 cm H₂OMin P (AVAPS minimum IPAP) 5 to 30 cm H₂OMin P (AVAPS minimum IPAP) 5 to 30 cm H₂OOxygen percentage 21 to 100 %Ramp time Off, 5 to 45 minRespiratory rate 4 to 60 beats per minute
Rise time 1 to 5Triggering and cycling Auto-adaptive (Auto-Trak) or Auto-Trak Plus (optional). Trigger: normal, +1 to +7. Cycle: -2, -1, normal, +1 to +6AVAPS target tidal volume 200 to 2000 ml BTPSMax E 0 to 100 cm H2O/lMax E 0 to 100 cm H₂O/lMax R 0 to 50 cm H2O/l/sMax R 0 to 50 cm H₂Ol/sPPV% 0 to 100 %Max P (PPV maximum pressure limit) 5 to 40 cm H₂OMax V (PPV maximum volume limit) 200 to 3500 ml
- Patient Data
-
Breath phase/trigger indicator Spontaneous, Timed, ExhalePIP 0 to 50 cm H₂OPIP 0 to 50 cm H₂OPatient/total leak 0 to 200 l/min BTPSPatient Trigger 0 to 100 %
Respiratory Rate 0 to 90 beats per minuteTi/Ttot 0 to 91 %Minute volume 0 to 99.0 l/min BTPSTidal volume 0 to 3000 ml BTPS
- Waveform window
-
Pressure waveform 0 - 50 cm H₂OFlow waveform 240-2401 /min BTPS
Volume waveform 0 - 3,500 ml BTPS
- Alarm adjustable range
-
Hi Rate (hi respiratory rate alarm) 5 - 90 beats per minuteLo Rate (low respiratory rate alarm) 1 - 89 beats per minuteHi Vr (high tidal volume alarm) 200 - 3,500 mlLo Vr (low tidal volume alarm) OFF - 1,500 mlHIP (high inspiratory pressure alarm) 5 - 50 cm H₂O
LIP (low inspiratory pressure alarm) OFF, 1 - 40 cm H₂OLo VE (low minute ventilation alarm) OFF, 0.1 - 991 /minLIP T (low inspiratory pressure delay time) 5 - 60 sec